*** MIDAS logfile (05SEPpl1.2) *** Th, 16 Feb 2006 18:26:51 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 16 Feb 2006 18:27:50 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 060216 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 346.29 stddev = 5.63 FIERA temperature = 290.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_060216 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1608.4 1609.4 1305.5 -0.09 -0.07 1.23 2 2.12 3186.2 3183.8 2883.8 0.18 0.13 1.10 3 4.25 6314.9 6313.2 5054.4 -0.08 -0.09 1.25 4 6.38 9453.6 9455.4 7498.8 -0.05 -0.04 1.26 5 9.57 14164.5 14165.9 11149.0 -0.02 -0.01 1.27 6 12.76 18889.6 18890.3 15050.5 0.07 0.08 1.26 7 15.95 23601.7 23599.5 18700.4 0.08 0.08 1.26 8 21.28 31451.7 31449.2 24550.9 0.06 0.06 1.28 9 26.59 39274.0 39275.0 30199.0 -0.01 -0.00 1.30 10 31.91 47053.8 47053.6 35628.9 -0.15 -0.15 1.32 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2122 to 62056 e- Average count rate measured: 1475 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.32 +/- 0.01 e-/ADU RON = 7.42 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.031 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay