*** MIDAS logfile (05SEPpl1.2) *** Mo, 13 Feb 2006 20:15:33 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 13 Feb 2006 20:16:30 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 060213 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 345.37 stddev = 5.93 FIERA temperature = 289.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_060213 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 804.2 804.9 722.4 0.05 0.09 1.11 2 2.00 1589.2 1589.8 1299.0 -0.10 -0.07 1.22 3 4.00 3162.3 3162.9 2537.3 -0.09 -0.09 1.25 4 6.00 4736.9 4737.3 3851.4 -0.05 -0.05 1.23 5 9.00 7097.4 7097.7 5728.8 -0.04 -0.05 1.24 6 12.00 9460.1 9461.1 7560.9 -0.00 -0.01 1.25 7 15.00 11821.5 11822.2 9393.1 0.01 -0.00 1.26 8 20.00 15762.9 15763.6 12654.2 0.05 0.04 1.25 9 25.00 19705.0 19704.1 15666.9 0.08 0.06 1.26 10 30.00 23645.7 23645.8 18587.3 0.09 0.08 1.27 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1021 to 30018 e- Average count rate measured: 787 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.27 +/- 0.01 e-/ADU RON = 7.53 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.025 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay