*** MIDAS logfile (05SEPpl1.2) *** Fr, 27 Jan 2006 15:33:31 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 27 Jan 2006 15:34:24 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 060127 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 331.22 stddev = 4.64 FIERA temperature = 291.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_060127 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 805.4 805.1 657.7 0.02 0.04 1.22 2 2.00 1594.7 1594.7 1286.5 -0.03 -0.03 1.24 3 4.00 3170.8 3170.7 2517.1 -0.09 -0.10 1.26 4 6.00 4748.6 4748.8 3791.1 -0.07 -0.07 1.25 5 9.00 7117.2 7116.5 5654.7 -0.04 -0.05 1.26 6 12.00 9485.6 9486.0 7927.9 -0.02 -0.02 1.20 7 15.00 11855.8 11856.3 9422.9 0.00 0.01 1.26 8 20.00 15808.6 15808.3 13286.4 0.05 0.05 1.19 9 25.00 19762.1 19763.0 15633.1 0.08 0.08 1.26 10 30.00 23718.9 23719.1 18629.0 0.11 0.11 1.27 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1013 to 29843 e- Average count rate measured: 789 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.26 +/- 0.02 e-/ADU RON = 5.84 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.024 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.10 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay