*** MIDAS logfile (05SEPpl1.2) *** Sa, 21 Jan 2006 16:17:11 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 21 Jan 2006 16:18:13 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 060121 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 334.12 stddev = 6.15 FIERA temperature = 291.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_060121 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 774.5 774.2 695.7 0.02 0.01 1.11 2 2.00 1553.6 1554.3 1300.4 -0.09 -0.05 1.19 3 4.00 3114.5 3113.6 2496.2 -0.04 -0.05 1.25 4 6.00 4673.6 4673.0 3771.4 -0.05 -0.07 1.24 5 9.00 7014.3 7014.1 5741.7 -0.04 -0.03 1.22 6 12.00 9355.5 9355.1 7495.1 -0.02 -0.02 1.25 7 15.00 11698.6 11699.2 9383.9 0.01 0.02 1.25 8 20.00 15604.4 15604.1 12232.7 0.04 0.04 1.28 9 25.00 19512.7 19512.2 15389.3 0.07 0.07 1.27 10 30.00 23422.3 23421.7 18708.0 0.09 0.09 1.25 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 981 to 29665 e- Average count rate measured: 780 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.27 +/- 0.01 e-/ADU RON = 7.79 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.08 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay