*** MIDAS logfile (05SEPpl1.2) *** Th, 19 Jan 2006 22:48:03 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 19 Jan 2006 22:48:57 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 060119 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.29 stddev = 7.45 FIERA temperature = 291.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_060119 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 771.6 771.3 760.0 0.05 0.04 1.02 2 2.00 1550.9 1550.5 1276.7 -0.03 -0.05 1.21 3 4.00 3108.2 3108.3 2517.1 -0.11 -0.10 1.23 4 6.00 4667.3 4667.3 3742.5 -0.09 -0.08 1.25 5 9.00 7007.7 7007.8 5654.6 -0.05 -0.04 1.24 6 12.00 9348.4 9348.3 8593.7 -0.02 -0.02 1.09 7 15.00 11690.8 11690.7 9617.9 0.01 0.01 1.22 8 20.00 15596.7 15597.3 12369.0 0.05 0.05 1.26 9 25.00 19502.4 19503.0 15558.9 0.08 0.08 1.25 10 30.00 23411.5 23412.1 18733.8 0.10 0.11 1.25 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 970 to 29451 e- Average count rate measured: 780 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.26 +/- 0.03 e-/ADU RON = 9.37 +/- 0.19 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.008 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.10 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay