*** MIDAS logfile (05SEPpl1.2) *** Sa, 14 Jan 2006 19:47:10 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 14 Jan 2006 19:48:05 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 060114 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 320.60 stddev = 4.26 FIERA temperature = 291.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_060114 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 785.6 785.2 644.9 0.10 0.06 1.22 2 2.00 1566.3 1566.7 1260.8 -0.12 -0.10 1.24 3 4.00 3131.9 3132.1 2502.1 -0.07 -0.06 1.25 4 6.00 4696.7 4696.7 3756.7 -0.07 -0.06 1.25 5 9.00 7044.4 7042.9 5550.7 -0.05 -0.07 1.27 6 12.00 9393.1 9393.7 7419.6 -0.03 -0.02 1.27 7 15.00 11745.3 11745.4 9379.9 0.01 0.01 1.25 8 20.00 15664.5 15664.4 12992.8 0.04 0.04 1.21 9 25.00 19586.5 19585.7 15738.5 0.08 0.07 1.24 10 30.00 23508.5 23510.6 18500.7 0.10 0.11 1.27 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 983 to 29438 e- Average count rate measured: 783 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.25 +/- 0.01 e-/ADU RON = 5.33 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay