*** MIDAS logfile (03SEPpl1.2) *** Fr, 04 Nov 2005 18:27:15 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 04 Nov 2005 18:28:20 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 051104 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 347.71 stddev = 4.09 FIERA temperature = 291.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_051104 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 818.4 818.5 728.4 0.03 0.07 1.12 2 2.00 1618.0 1617.5 1290.9 -0.06 -0.06 1.25 3 4.00 3217.4 3217.5 2563.1 -0.09 -0.07 1.26 4 6.00 4817.6 4817.3 3906.8 -0.08 -0.08 1.23 5 9.00 7219.4 7219.7 5723.5 -0.05 -0.05 1.26 6 12.00 9622.3 9620.6 7622.0 -0.02 -0.04 1.26 7 15.00 12026.5 12027.2 9673.0 0.00 0.00 1.24 8 20.00 16037.8 16037.1 12668.3 0.06 0.05 1.27 9 25.00 20048.1 20049.3 15844.2 0.09 0.08 1.27 10 30.00 24060.7 24059.7 19343.2 0.12 0.10 1.24 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1028 to 30219 e- Average count rate measured: 801 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.26 +/- 0.01 e-/ADU RON = 5.13 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.025 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.08 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay