*** MIDAS logfile (03SEPpl1.2) *** Th, 20 Oct 2005 16:41:52 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Th, 20 Oct 2005 16:42:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 051020 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 347.66 stddev = 4.10 FIERA temperature = 291.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_051020 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 820.4 820.6 667.6 0.07 0.08 1.23 2 2.00 1621.1 1621.1 1299.1 -0.07 -0.07 1.25 3 4.00 3224.1 3224.6 2644.2 -0.08 -0.07 1.22 4 6.00 4826.8 4827.5 3852.7 -0.09 -0.08 1.25 5 9.00 7233.4 7234.3 5757.9 -0.05 -0.05 1.26 6 12.00 9641.6 9641.3 7708.6 -0.03 -0.03 1.25 7 15.00 12051.0 12051.2 10002.8 0.00 0.00 1.20 8 20.00 16071.2 16070.4 13066.2 0.06 0.05 1.23 9 25.00 20088.8 20089.2 16054.4 0.08 0.08 1.25 10 30.00 24109.3 24110.5 18765.9 0.11 0.11 1.28 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1039 to 30531 e- Average count rate measured: 802 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.27 +/- 0.01 e-/ADU RON = 5.19 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.025 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.08 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay