*** MIDAS logfile (03SEPpl1.2) *** We, 07 Sep 2005 15:37:44 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** We, 07 Sep 2005 15:38:42 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 050907 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 347.22 stddev = 5.20 FIERA temperature = 290.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_050907 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 828.4 828.8 678.0 0.07 0.03 1.22 2 2.00 1631.9 1632.1 1325.2 -0.05 -0.08 1.23 3 4.00 3240.9 3240.5 2598.3 -0.07 -0.09 1.25 4 6.00 4848.6 4849.2 3854.6 -0.09 -0.08 1.26 5 9.00 7263.7 7264.2 5744.7 -0.05 -0.04 1.26 6 12.00 9678.4 9678.4 7675.1 -0.04 -0.03 1.26 7 15.00 12097.9 12097.6 9559.7 0.01 0.02 1.27 8 20.00 16129.1 16129.3 12860.6 0.05 0.06 1.25 9 25.00 20161.0 20162.0 16117.5 0.08 0.09 1.25 10 30.00 24195.3 24194.5 19136.2 0.11 0.12 1.26 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1044 to 30490 e- Average count rate measured: 805 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.26 +/- 0.00 e-/ADU RON = 6.55 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.031 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay