*** MIDAS logfile (03SEPpl1.2) *** Sa, 05 Feb 2005 16:02:53 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 05 Feb 2005 16:03:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 050205 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 332.03 stddev = 5.83 FIERA temperature = 291.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_050205 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 842.2 842.4 692.0 0.09 0.08 1.22 2 2.00 1666.0 1665.6 1337.0 -0.05 -0.06 1.25 3 4.00 3313.1 3313.7 2648.1 -0.11 -0.10 1.25 4 6.00 4961.4 4961.1 3917.3 -0.10 -0.10 1.27 5 9.00 7435.0 7435.0 5933.8 -0.08 -0.07 1.25 6 12.00 9911.1 9911.6 7852.0 -0.04 -0.03 1.26 7 15.00 12388.9 12388.9 10065.1 -0.00 -0.00 1.23 8 20.00 16522.4 16521.5 12935.9 0.06 0.05 1.28 9 25.00 20655.8 20655.9 16284.4 0.09 0.09 1.27 10 30.00 24792.6 24794.0 19473.5 0.13 0.14 1.27 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1073 to 31577 e- Average count rate measured: 825 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.27 +/- 0.01 e-/ADU RON = 7.43 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.023 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.10 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay