*** MIDAS logfile (03SEPpl1.2) *** Su, 30 Jan 2005 14:51:14 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 30 Jan 2005 14:52:18 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 050130 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 331.61 stddev = 5.87 FIERA temperature = 290.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_050130 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 857.2 857.0 781.2 0.04 0.08 1.10 2 2.00 1683.8 1684.6 1366.0 -0.07 -0.01 1.23 3 4.00 3337.0 3337.2 2648.8 -0.10 -0.10 1.26 4 6.00 4991.2 4991.1 4015.6 -0.09 -0.10 1.24 5 9.00 7473.3 7473.4 5909.9 -0.06 -0.08 1.26 6 12.00 9957.9 9958.1 7885.7 -0.03 -0.04 1.26 7 15.00 12443.7 12444.4 9833.3 0.00 -0.01 1.27 8 20.00 16591.8 16591.6 13023.2 0.06 0.04 1.27 9 25.00 20740.5 20741.2 16911.6 0.11 0.09 1.23 10 30.00 24891.8 24892.2 19323.7 0.14 0.12 1.29 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1089 to 31625 e- Average count rate measured: 828 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.27 +/- 0.01 e-/ADU RON = 7.46 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.037 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.10 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay