*** MIDAS logfile (05SEPpl1.2) *** Su, 02 Mar 2008 16:19:46 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 02 Mar 2008 16:20:04 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD1 TAKEN ON 080302 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 243.06 stddev = 5.59 FIERA temperature = 288.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD1_080302 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1299.2 1299.5 1041.5 0.06 0.08 1.25 2 8.51 2601.7 2601.5 2290.3 -0.04 -0.05 1.14 3 17.02 5204.2 5204.7 4033.3 -0.13 -0.12 1.29 4 25.53 7810.5 7809.8 6263.5 -0.11 -0.12 1.25 5 38.29 11725.3 11725.1 9061.2 -0.05 -0.06 1.29 6 51.06 15640.5 15640.8 11824.0 -0.02 -0.02 1.32 7 63.83 19562.4 19561.9 14573.1 0.03 0.03 1.34 8 85.10 26099.6 26100.6 19088.4 0.09 0.09 1.37 9 106.38 32628.1 32629.0 23358.8 0.10 0.10 1.40 10 127.66 39144.8 39144.5 27582.5 0.07 0.07 1.42 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1856 to 55906 e- Average count rate measured: 306 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.43 +/- 0.02 e-/ADU RON = 7.99 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.014 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay