*** MIDAS logfile (05SEPpl1.2) *** Sa, 09 Feb 2008 15:49:09 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 09 Feb 2008 15:49:26 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD1 TAKEN ON 080209 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 267.14 stddev = 4.70 FIERA temperature = 288.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD1_080209 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1304.0 1304.5 1034.8 0.07 0.08 1.26 2 8.51 2611.5 2611.5 2045.8 -0.04 -0.05 1.28 3 17.02 5224.4 5224.7 4046.1 -0.12 -0.12 1.29 4 25.53 7840.4 7840.4 6046.6 -0.11 -0.11 1.30 5 38.30 11770.1 11770.5 9021.6 -0.05 -0.05 1.30 6 51.06 15700.2 15700.1 11777.8 -0.02 -0.02 1.33 7 63.83 19636.0 19636.9 14767.6 0.03 0.03 1.33 8 85.10 26198.0 26198.2 19366.5 0.08 0.08 1.35 9 106.38 32752.5 32752.2 23421.8 0.09 0.09 1.40 10 127.66 39295.3 39292.8 40194.2 0.07 0.07 0.98 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1460 to 43989 e- Average count rate measured: 308 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.12 +/- 0.10 e-/ADU RON = 5.26 +/- 0.46 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.016 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay