*** MIDAS logfile (05SEPpl1.2) *** Th, 10 Jan 2008 14:59:31 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 10 Jan 2008 14:59:49 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD1 TAKEN ON 080110 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 268.20 stddev = 4.85 FIERA temperature = 288.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD1_080110 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1349.0 1349.4 1123.5 0.07 0.08 1.20 2 8.51 2694.8 2694.4 2110.0 -0.04 -0.05 1.28 3 17.02 5383.8 5383.8 4196.8 -0.12 -0.12 1.28 4 25.53 8075.5 8075.8 6295.3 -0.12 -0.11 1.28 5 38.29 12120.9 12121.0 9186.7 -0.05 -0.05 1.32 6 51.06 16165.8 16166.1 12142.6 -0.02 -0.02 1.33 7 63.83 20218.1 20218.9 15023.1 0.04 0.04 1.35 8 85.10 26971.7 26971.0 19668.1 0.09 0.09 1.37 9 106.38 33715.2 33715.7 24103.7 0.09 0.10 1.40 10 127.66 40441.5 40440.9 28170.9 0.05 0.05 1.44 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1933 to 57954 e- Average count rate measured: 317 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.43 +/- 0.02 e-/ADU RON = 6.95 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay