*** MIDAS logfile (05SEPpl1.2) *** We, 12 Dec 2007 14:06:21 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 12 Dec 2007 14:06:38 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD1 TAKEN ON 071212 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 265.70 stddev = 4.71 FIERA temperature = 283.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD1_071212 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1320.5 1320.8 1071.3 0.06 0.06 1.23 2 8.51 2644.8 2644.6 2082.7 -0.04 -0.06 1.27 3 17.02 5291.3 5291.6 4151.0 -0.11 -0.11 1.27 4 25.53 7940.0 7940.3 6105.6 -0.12 -0.11 1.30 5 38.29 11921.1 11921.4 9084.5 -0.05 -0.05 1.31 6 51.06 15901.7 15902.5 11929.0 -0.02 -0.01 1.33 7 63.83 19889.9 19889.5 14784.8 0.04 0.04 1.35 8 85.10 26536.0 26536.1 19449.9 0.09 0.09 1.36 9 106.38 33172.3 33173.0 23611.5 0.09 0.10 1.40 10 127.66 39794.7 39794.4 27856.0 0.06 0.06 1.43 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1887 to 56875 e- Average count rate measured: 312 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.43 +/- 0.02 e-/ADU RON = 6.72 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.016 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay