*** MIDAS logfile (05SEPpl1.2) *** We, 19 Sep 2007 17:55:33 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 19 Sep 2007 17:55:51 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD1 TAKEN ON 070919 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 314.82 stddev = 4.66 FIERA temperature = 282.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD1_070919 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1405.5 1405.5 1139.5 -0.01 0.01 1.23 2 8.51 2802.6 2802.3 2251.9 0.01 0.01 1.24 3 17.02 5593.0 5592.7 4404.0 -0.03 -0.04 1.27 4 25.53 8384.9 8385.4 6530.5 -0.02 -0.02 1.28 5 38.29 12577.5 12577.9 9708.8 0.01 0.01 1.30 6 51.06 16767.9 16768.2 12839.4 0.02 0.02 1.31 7 63.83 20962.2 20962.4 15834.5 0.05 0.04 1.32 8 85.10 27948.1 27948.9 20577.5 0.05 0.05 1.36 9 106.38 34915.4 34915.7 25209.0 0.00 -0.00 1.39 10 127.66 41860.1 41860.8 29543.5 -0.08 -0.09 1.42 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1990 to 59257 e- Average count rate measured: 328 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.42 +/- 0.02 e-/ADU RON = 6.60 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.033 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.05/-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay