*** MIDAS logfile (05SEPpl1.2) *** Su, 01 Jul 2007 19:04:52 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 01 Jul 2007 19:05:09 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD1 TAKEN ON 070701 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 322.45 stddev = 5.69 FIERA temperature = 284.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD1_070701 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1431.5 1431.9 1160.8 -0.01 -0.01 1.23 2 8.51 2855.3 2855.3 2282.5 0.02 0.02 1.25 3 17.02 5698.2 5698.1 4506.5 -0.02 -0.03 1.26 4 25.53 8542.8 8542.5 6678.5 -0.03 -0.03 1.28 5 38.30 12815.1 12814.9 9964.3 0.02 0.02 1.29 6 51.06 17083.9 17083.8 12957.3 0.02 0.02 1.32 7 63.83 21357.7 21357.8 16037.2 0.05 0.05 1.33 8 85.10 28475.4 28475.9 21033.0 0.05 0.05 1.35 9 106.38 35573.1 35573.0 25549.7 -0.00 -0.00 1.39 10 127.66 42646.4 42646.3 31139.7 -0.09 -0.09 1.37 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1990 to 59281 e- Average count rate measured: 334 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.39 +/- 0.01 e-/ADU RON = 7.90 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.031 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.05/-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay