*** MIDAS logfile (05SEPpl1.2) *** Sa, 28 Apr 2007 17:47:15 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 28 Apr 2007 17:47:33 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD1 TAKEN ON 070428 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 323.59 stddev = 6.06 FIERA temperature = 283.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD1_070428 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1398.6 1398.6 1144.2 0.01 0.00 1.22 2 8.51 2803.7 2803.3 2244.5 0.01 -0.01 1.25 3 17.02 5611.1 5611.2 4436.2 -0.03 -0.03 1.26 4 25.53 8419.7 8420.4 6690.9 -0.04 -0.03 1.26 5 38.30 12639.1 12639.1 9766.8 0.01 0.02 1.29 6 51.06 16855.0 16854.8 12883.9 0.02 0.02 1.31 7 63.83 21075.6 21076.2 15830.0 0.04 0.05 1.33 8 85.10 28104.8 28105.5 20866.5 0.05 0.06 1.35 9 106.38 35116.7 35116.4 25262.6 0.01 0.01 1.39 10 127.66 42103.5 42103.1 29591.4 -0.08 -0.08 1.42 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1987 to 59820 e- Average count rate measured: 330 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.42 +/- 0.02 e-/ADU RON = 8.61 +/- 0.12 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.016 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.05/-0.08 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay