*** MIDAS logfile (05SEPpl1.2) *** Mo, 12 Mar 2007 16:38:32 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 12 Mar 2007 16:38:50 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD1 TAKEN ON 070312 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 323.48 stddev = 4.82 FIERA temperature = 282.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD1_070312 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1425.1 1425.0 1154.0 0.02 0.01 1.23 2 8.51 2851.2 2851.3 2274.3 -0.01 -0.01 1.25 3 17.02 5702.1 5702.2 4660.5 -0.03 -0.02 1.22 4 25.53 8553.4 8553.8 6671.6 -0.03 -0.02 1.28 5 38.30 12836.9 12836.0 9894.7 0.02 0.01 1.30 6 51.06 17116.6 17116.4 13136.2 0.02 0.02 1.30 7 63.83 21402.0 21401.7 16110.3 0.05 0.05 1.33 8 85.10 28537.3 28536.9 20890.2 0.05 0.05 1.37 9 106.38 35653.8 35653.1 25601.2 0.00 0.00 1.39 10 127.66 42743.7 42743.5 29959.8 -0.09 -0.09 1.43 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2035 to 61029 e- Average count rate measured: 335 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.43 +/- 0.02 e-/ADU RON = 6.88 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.05/-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay