*** MIDAS logfile (05SEPpl1.2) *** We, 03 Jan 2007 16:07:59 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 03 Jan 2007 16:08:17 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD1 TAKEN ON 070103 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 323.83 stddev = 4.81 FIERA temperature = 288.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD1_070103 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1434.8 1435.3 1170.3 0.01 0.01 1.23 2 8.51 2877.0 2876.8 2295.2 0.01 -0.01 1.25 3 17.02 5757.6 5757.9 4531.8 -0.02 -0.02 1.27 4 25.53 8639.8 8639.5 6770.7 -0.02 -0.03 1.28 5 38.30 12968.0 12967.8 10011.0 0.02 0.02 1.30 6 51.06 17293.4 17293.3 13241.2 0.02 0.02 1.31 7 63.83 21623.9 21624.9 16271.0 0.05 0.06 1.33 8 85.10 28836.0 28834.7 21126.0 0.06 0.05 1.36 9 106.38 36023.7 36023.0 25713.6 -0.01 -0.01 1.40 10 127.66 43189.1 43190.1 30287.4 -0.10 -0.10 1.43 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2050 to 61694 e- Average count rate measured: 339 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.43 +/- 0.02 e-/ADU RON = 6.87 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.015 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.06/-0.10 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay