*** MIDAS logfile (05SEPpl1.2) *** Sa, 04 Nov 2006 15:13:43 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 04 Nov 2006 15:14:56 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD1 TAKEN ON 061104 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 318.24 stddev = 4.58 FIERA temperature = 288.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD1_061104 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1490.4 1491.0 1231.3 0.01 0.01 1.21 2 8.51 2978.0 2978.4 2348.1 0.01 0.00 1.27 3 17.02 5949.1 5949.3 4648.4 -0.03 -0.04 1.28 4 25.53 8921.7 8921.8 6939.9 -0.04 -0.03 1.29 5 38.29 13386.4 13385.9 10382.7 0.01 0.01 1.29 6 51.06 17849.3 17849.0 13554.1 0.02 0.02 1.32 7 63.83 22317.0 22316.6 16650.4 0.05 0.05 1.34 8 85.10 29758.5 29757.8 21957.4 0.06 0.06 1.36 9 106.38 37174.9 37174.4 26445.6 0.00 0.00 1.41 10 127.66 44567.4 44566.9 30938.6 -0.09 -0.09 1.44 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2142 to 64025 e- Average count rate measured: 349 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.44 +/- 0.02 e-/ADU RON = 6.58 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.016 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.06/-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay