*** MIDAS logfile (05SEPpl1.2) *** We, 13 Sep 2006 15:52:40 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 13 Sep 2006 15:53:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD1 TAKEN ON 060913 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 258.68 stddev = 4.60 FIERA temperature = 287.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD1_060913 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1460.5 1460.8 1146.5 0.08 0.09 1.27 2 8.51 2915.4 2915.2 2262.7 -0.06 -0.07 1.29 3 17.02 5823.0 5823.3 4532.9 -0.15 -0.15 1.28 4 25.53 8734.7 8734.8 6678.6 -0.14 -0.14 1.31 5 38.30 13110.3 13110.5 9868.9 -0.06 -0.06 1.33 6 51.06 17486.7 17486.7 13127.3 -0.02 -0.02 1.33 7 63.83 21873.0 21872.2 16150.1 0.05 0.04 1.35 8 85.10 29183.6 29184.0 21186.5 0.12 0.12 1.38 9 106.38 36477.2 36476.6 25726.0 0.11 0.11 1.42 10 127.66 43754.7 43756.0 30037.3 0.07 0.08 1.46 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2119 to 63475 e- Average count rate measured: 342 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.45 +/- 0.02 e-/ADU RON = 6.67 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay