*** MIDAS logfile (05SEPpl1.2) *** Su, 20 Aug 2006 18:08:06 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 20 Aug 2006 18:09:25 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD1 TAKEN ON 060820 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 259.55 stddev = 4.70 FIERA temperature = 288.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD1_060820 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1432.3 1432.6 1126.5 0.08 0.08 1.27 2 8.51 2872.8 2873.1 2235.1 -0.05 -0.06 1.29 3 17.02 5752.4 5752.0 4431.0 -0.14 -0.15 1.30 4 25.53 8634.8 8635.3 6599.9 -0.14 -0.13 1.31 5 38.29 12966.7 12967.8 9746.8 -0.07 -0.06 1.33 6 51.06 17301.4 17301.4 12876.4 -0.02 -0.02 1.34 7 63.83 21643.4 21642.7 16012.8 0.04 0.04 1.35 8 85.10 28882.4 28883.0 20920.0 0.11 0.12 1.38 9 106.38 36105.8 36106.6 25764.6 0.11 0.12 1.40 10 127.66 43314.0 43313.6 29885.9 0.08 0.08 1.45 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2063 to 62371 e- Average count rate measured: 339 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.44 +/- 0.02 e-/ADU RON = 6.76 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.032 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay