*** MIDAS logfile (05SEPpl1.2) *** Sa, 01 Jul 2006 15:37:16 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 01 Jul 2006 15:38:32 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD1 TAKEN ON 060701 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 258.45 stddev = 4.71 FIERA temperature = 288.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD1_060701 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1464.0 1463.7 1147.5 0.09 0.09 1.28 2 8.51 2928.2 2927.9 2287.7 -0.08 -0.08 1.28 3 17.02 5856.2 5856.1 4485.2 -0.15 -0.15 1.31 4 25.53 8788.4 8788.4 6693.6 -0.14 -0.14 1.31 5 38.30 13195.1 13194.3 9957.9 -0.06 -0.07 1.33 6 51.06 17603.1 17602.9 13083.2 -0.01 -0.02 1.35 7 63.83 22019.3 22019.7 16230.7 0.05 0.05 1.36 8 85.10 29380.9 29381.2 21059.9 0.12 0.12 1.40 9 106.38 36726.5 36727.4 25846.6 0.12 0.12 1.42 10 127.66 44054.4 44054.9 30375.2 0.08 0.07 1.45 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2123 to 63900 e- Average count rate measured: 345 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.45 +/- 0.02 e-/ADU RON = 6.84 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.013 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay