*** MIDAS logfile (05SEPpl1.2) *** Fr, 27 Jan 2006 13:51:52 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 27 Jan 2006 13:53:03 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD1 TAKEN ON 060127 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 285.43 stddev = 4.39 FIERA temperature = 291.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD1_060127 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 790.2 790.5 641.6 0.13 0.11 1.23 2 8.00 1572.1 1572.4 1290.3 -0.19 -0.19 1.22 3 16.00 3144.9 3144.7 2474.2 -0.05 -0.06 1.27 4 24.00 4715.4 4715.8 3696.1 -0.06 -0.05 1.28 5 36.00 7071.6 7071.8 5523.3 -0.05 -0.05 1.28 6 48.00 9431.3 9431.2 7350.6 -0.01 -0.01 1.28 7 60.00 11791.5 11791.2 9032.4 0.01 0.02 1.31 8 80.00 15726.7 15724.1 11988.3 0.05 0.04 1.31 9 100.00 19663.3 19663.3 14757.4 0.08 0.09 1.33 10 120.00 23600.8 23600.2 17587.1 0.10 0.11 1.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1063 to 31736 e- Average count rate measured: 196 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.34 +/- 0.01 e-/ADU RON = 5.90 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.022 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay