*** MIDAS logfile (05SEPpl1.2) *** Sa, 21 Jan 2006 15:36:45 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 21 Jan 2006 15:38:05 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD1 TAKEN ON 060121 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 328.59 stddev = 7.34 FIERA temperature = 291.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD1_060121 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 773.7 773.6 677.7 0.08 0.05 1.14 2 8.00 1548.5 1548.6 1400.6 -0.06 -0.05 1.11 3 16.00 3099.1 3099.4 2498.2 -0.09 -0.08 1.24 4 24.00 4650.6 4650.7 3667.0 -0.09 -0.08 1.27 5 36.00 6979.4 6979.5 5484.9 -0.06 -0.05 1.27 6 48.00 9309.9 9309.5 7236.9 -0.03 -0.03 1.29 7 60.00 11641.8 11641.2 9027.0 0.01 0.00 1.29 8 80.00 15529.7 15529.5 12139.5 0.05 0.05 1.28 9 100.00 19419.6 19419.9 14707.0 0.08 0.09 1.32 10 120.00 23310.8 23310.7 17460.1 0.11 0.11 1.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1034 to 31157 e- Average count rate measured: 194 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.34 +/- 0.01 e-/ADU RON = 9.81 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.014 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay