*** MIDAS logfile (05SEPpl1.2) *** Sa, 08 Dec 2007 16:19:44 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 08 Dec 2007 16:19:59 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 071208 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 263.07 stddev = 5.51 FIERA temperature = 286.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_071208 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1643.5 1643.7 1353.8 -0.03 -0.02 1.21 2 2.12 3178.8 3179.1 2587.1 0.18 0.16 1.23 3 4.25 6223.1 6221.0 5059.9 -0.14 -0.16 1.23 4 6.38 9281.9 9282.5 7511.8 -0.09 -0.09 1.24 5 9.57 13868.7 13871.1 11160.9 -0.06 -0.05 1.24 6 12.76 18473.3 18473.9 14949.4 0.04 0.04 1.24 7 15.95 23065.0 23065.7 18415.8 0.05 0.06 1.25 8 21.27 30719.9 30720.1 24307.0 0.07 0.07 1.26 9 26.59 38361.4 38362.1 30468.9 0.04 0.05 1.26 10 31.91 45962.3 45963.0 35621.6 -0.06 -0.06 1.29 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2108 to 58937 e- Average count rate measured: 1437 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.28 +/- 0.01 e-/ADU RON = 7.06 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.084 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay