*** MIDAS logfile (05SEPpl1.2) *** We, 14 Nov 2007 17:11:44 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 14 Nov 2007 17:11:59 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 071114 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 263.84 stddev = 5.04 FIERA temperature = 282.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_071114 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1662.7 1661.8 1362.9 -0.03 -0.07 1.22 2 2.12 3203.5 3204.2 2599.0 0.19 0.21 1.23 3 4.25 6255.3 6255.5 5094.0 -0.14 -0.14 1.23 4 6.38 9323.2 9323.4 7486.8 -0.09 -0.08 1.25 5 9.57 13921.8 13922.6 11246.8 -0.06 -0.06 1.24 6 12.76 18539.9 18539.8 14914.6 0.04 0.04 1.24 7 15.95 23142.8 23143.3 18633.7 0.05 0.05 1.24 8 21.27 30818.5 30817.6 25362.0 0.06 0.06 1.22 9 26.59 38481.2 38481.6 30647.9 0.04 0.04 1.26 10 31.91 46104.4 46104.5 36217.1 -0.06 -0.06 1.27 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2097 to 58152 e- Average count rate measured: 1441 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.26 +/- 0.01 e-/ADU RON = 6.35 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.095 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.20/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay