*** MIDAS logfile (05SEPpl1.2) *** Th, 18 Oct 2007 17:11:01 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 18 Oct 2007 17:11:16 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 071018 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 308.51 stddev = 5.00 FIERA temperature = 286.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_071018 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1660.3 1660.7 1386.9 -0.09 -0.08 1.20 2 2.12 3212.4 3211.6 3593.7 0.24 0.21 0.89 3 4.25 6283.6 6282.4 5160.6 -0.09 -0.10 1.22 4 6.38 9371.1 9371.8 7694.0 -0.04 -0.04 1.22 5 9.57 13999.2 14000.7 12288.3 -0.01 -0.01 1.14 6 12.76 18641.3 18642.4 15146.7 0.06 0.07 1.23 7 15.95 23271.3 23272.4 18826.1 0.06 0.07 1.24 8 21.27 30981.8 30982.9 24798.3 0.04 0.04 1.25 9 26.59 38677.6 38678.8 30626.7 -0.02 -0.01 1.26 10 31.91 46327.8 46330.1 36177.0 -0.15 -0.14 1.28 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2146 to 59889 e- Average count rate measured: 1450 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.29 +/- 0.02 e-/ADU RON = 6.47 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.087 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.22/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay