*** MIDAS logfile (05SEPpl1.2) *** Su, 14 Oct 2007 16:12:21 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 14 Oct 2007 16:12:35 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 071014 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 315.72 stddev = 5.34 FIERA temperature = 285.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_071014 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1695.5 1695.3 1425.1 -0.12 -0.09 1.19 2 2.12 3269.3 3268.6 2704.9 0.22 0.24 1.21 3 4.25 6382.1 6381.4 5231.6 -0.09 -0.09 1.22 4 6.38 9512.9 9511.4 7789.6 -0.03 -0.04 1.22 5 9.57 14203.7 14201.2 11655.3 -0.01 -0.02 1.22 6 12.76 18907.9 18907.3 15252.3 0.07 0.06 1.24 7 15.95 23598.0 23598.3 19282.9 0.06 0.06 1.22 8 21.27 31415.9 31415.6 25237.0 0.04 0.04 1.24 9 26.59 39212.9 39214.4 31286.7 -0.01 -0.02 1.25 10 31.91 46962.9 46962.9 36658.8 -0.15 -0.16 1.28 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2158 to 59784 e- Average count rate measured: 1470 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.27 +/- 0.01 e-/ADU RON = 6.79 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.095 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.23/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay