*** MIDAS logfile (05SEPpl1.2) *** Th, 27 Sep 2007 18:24:38 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 27 Sep 2007 18:24:53 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 070927 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 319.98 stddev = 5.50 FIERA temperature = 282.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_070927 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1625.2 1625.9 1359.7 -0.09 -0.08 1.20 2 2.12 3172.1 3172.1 2622.7 0.20 0.21 1.21 3 4.25 6238.9 6239.5 5126.3 -0.10 -0.09 1.22 4 6.38 9321.6 9321.1 7688.0 -0.04 -0.04 1.21 5 9.57 13941.3 13941.2 11489.0 -0.01 -0.02 1.21 6 12.76 18575.5 18577.4 15190.2 0.07 0.07 1.22 7 15.95 23196.8 23194.2 18595.9 0.07 0.06 1.25 8 21.27 30896.2 30896.8 26817.0 0.05 0.05 1.15 9 26.59 38576.6 38576.7 30573.8 -0.01 -0.01 1.26 10 31.91 46213.3 46215.2 36475.8 -0.14 -0.14 1.27 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2035 to 57856 e- Average count rate measured: 1447 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.25 +/- 0.03 e-/ADU RON = 6.88 +/- 0.14 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.064 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.20/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay