*** MIDAS logfile (05SEPpl1.2) *** We, 19 Sep 2007 18:24:49 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 19 Sep 2007 18:25:04 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 070919 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 308.62 stddev = 4.98 FIERA temperature = 282.3 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_070919 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1695.8 1697.2 1403.5 -0.15 -0.11 1.21 2 2.12 3274.8 3274.8 2704.8 0.30 0.29 1.21 3 4.25 6390.4 6390.2 5272.6 -0.07 -0.08 1.21 4 6.38 9521.9 9521.6 8026.8 -0.03 -0.04 1.19 5 9.57 14218.3 14217.8 11528.0 -0.01 -0.01 1.23 6 12.76 18927.9 18927.0 15666.2 0.07 0.06 1.21 7 15.95 23622.3 23622.0 19003.9 0.06 0.05 1.24 8 21.27 31445.7 31446.6 25157.8 0.03 0.03 1.25 9 26.59 39249.2 39250.8 34876.1 -0.03 -0.03 1.13 10 31.91 47008.7 47010.1 37778.9 -0.16 -0.16 1.24 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2046 to 56683 e- Average count rate measured: 1471 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.21 +/- 0.03 e-/ADU RON = 6.01 +/- 0.16 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.096 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.30/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay