*** MIDAS logfile (05SEPpl1.2) *** We, 05 Sep 2007 17:02:35 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 05 Sep 2007 17:02:50 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 070905 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 313.11 stddev = 5.18 FIERA temperature = 283.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_070905 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1634.8 1633.9 1370.9 -0.09 -0.05 1.19 2 2.12 3193.4 3193.3 2628.0 0.19 0.22 1.22 3 4.25 6280.5 6279.5 5170.6 -0.09 -0.10 1.21 4 6.38 9383.1 9383.6 7651.9 -0.04 -0.04 1.23 5 9.57 14035.3 14032.4 11479.6 -0.01 -0.03 1.22 6 12.76 18698.6 18699.3 15190.6 0.07 0.06 1.23 7 15.95 23351.1 23351.6 18745.2 0.07 0.06 1.25 8 21.27 31100.2 31100.4 24926.6 0.05 0.04 1.25 9 26.59 38831.4 38832.2 31038.4 -0.01 -0.02 1.25 10 31.91 46519.0 46519.4 37011.2 -0.14 -0.15 1.26 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2058 to 58572 e- Average count rate measured: 1457 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.26 +/- 0.00 e-/ADU RON = 6.53 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.063 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.20/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay