*** MIDAS logfile (05SEPpl1.2) *** Mo, 27 Aug 2007 15:53:06 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 27 Aug 2007 15:53:21 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 070827 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 308.58 stddev = 5.00 FIERA temperature = 286.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_070827 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1613.7 1614.9 1344.2 -0.10 -0.07 1.20 2 2.12 3172.7 3172.9 2612.2 0.21 0.22 1.21 3 4.25 6257.1 6257.5 5158.3 -0.11 -0.11 1.21 4 6.38 9359.4 9359.4 7621.8 -0.04 -0.05 1.23 5 9.57 14010.5 14009.3 11365.4 -0.01 -0.02 1.23 6 12.76 18673.5 18674.3 15312.8 0.07 0.07 1.22 7 15.95 23323.9 23323.1 18678.7 0.07 0.07 1.25 8 21.27 31070.0 31071.0 24861.8 0.05 0.05 1.25 9 26.59 38797.5 38798.6 30946.9 -0.01 -0.01 1.25 10 31.91 46484.0 46481.7 36347.5 -0.14 -0.14 1.28 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2055 to 59171 e- Average count rate measured: 1456 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.27 +/- 0.01 e-/ADU RON = 6.37 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.050 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.21/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay