*** MIDAS logfile (05SEPpl1.2) *** We, 15 Nov 2006 15:22:08 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 15 Nov 2006 15:22:22 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 061115 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 320.75 stddev = 5.26 FIERA temperature = 288.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_061115 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1794.7 1795.7 1494.5 -0.08 -0.06 1.20 2 2.13 3472.6 3472.0 2861.0 0.22 0.20 1.21 3 4.25 6793.4 6792.9 6187.1 -0.08 -0.09 1.10 4 6.38 10130.0 10128.7 8227.7 -0.04 -0.04 1.23 5 9.57 15130.7 15131.7 12430.4 -0.02 -0.02 1.22 6 12.76 20150.9 20151.5 16783.6 0.07 0.07 1.20 7 15.96 25155.1 25154.7 21088.1 0.07 0.07 1.19 8 21.28 33491.4 33492.0 27642.9 0.05 0.05 1.21 9 26.59 41802.5 41801.5 32792.3 -0.02 -0.02 1.27 10 31.91 50058.3 50058.1 39071.1 -0.17 -0.17 1.28 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2295 to 63987 e- Average count rate measured: 1567 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.28 +/- 0.02 e-/ADU RON = 6.73 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.085 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.21/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay