*** MIDAS logfile (05SEPpl1.2) *** Sa, 04 Nov 2006 12:50:54 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 04 Nov 2006 12:51:52 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 061104 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 322.80 stddev = 4.27 FIERA temperature = 287.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_061104 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.05 1786.7 1789.5 1477.7 -0.08 -0.05 1.21 2 2.12 3470.0 3468.8 2843.6 0.15 0.17 1.22 3 4.25 6796.0 6795.0 5681.0 -0.09 -0.10 1.20 4 6.38 10134.7 10136.0 8176.8 -0.05 -0.04 1.24 5 9.57 15144.4 15145.1 12728.6 -0.02 -0.02 1.19 6 12.76 20171.8 20172.8 16388.7 0.09 0.08 1.23 7 15.95 25184.7 25182.2 20415.5 0.09 0.08 1.23 8 21.27 33529.5 33530.1 26850.0 0.07 0.06 1.25 9 26.59 41848.7 41848.9 32916.7 -0.00 -0.01 1.27 10 31.91 50119.2 50122.6 38949.9 -0.15 -0.16 1.29 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2297 to 64392 e- Average count rate measured: 1569 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.28 +/- 0.01 e-/ADU RON = 5.49 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.084 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay