*** MIDAS logfile (05SEPpl1.2) *** We, 18 Oct 2006 15:19:57 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 18 Oct 2006 15:20:59 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 061018 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 322.52 stddev = 4.94 FIERA temperature = 288.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_061018 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1710.2 1710.6 1465.2 -0.07 -0.06 1.17 2 2.12 3364.3 3363.0 2762.4 0.22 0.19 1.22 3 4.25 6636.9 6637.1 5484.5 -0.11 -0.11 1.21 4 6.38 9928.2 9928.6 8203.9 -0.06 -0.05 1.21 5 9.57 14862.4 14863.7 12773.5 -0.02 -0.02 1.16 6 12.76 19812.8 19814.7 16097.5 0.07 0.07 1.23 7 15.95 24749.6 24750.3 20042.6 0.07 0.07 1.23 8 21.27 32972.0 32975.6 26101.1 0.05 0.06 1.26 9 26.59 41175.6 41175.1 32845.5 -0.00 -0.01 1.25 10 31.91 49321.9 49321.5 38414.3 -0.16 -0.16 1.28 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2192 to 63217 e- Average count rate measured: 1546 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.28 +/- 0.01 e-/ADU RON = 6.33 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.047 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.21/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay