*** MIDAS logfile (05SEPpl1.2) *** Fr, 06 Oct 2006 14:42:37 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 06 Oct 2006 14:43:29 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 061006 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 321.69 stddev = 4.68 FIERA temperature = 288.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_061006 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1802.2 1801.8 1494.9 -0.09 -0.09 1.21 2 2.12 3489.7 3489.6 2864.7 0.18 0.18 1.22 3 4.25 6835.4 6833.8 5568.1 -0.08 -0.09 1.23 4 6.38 10195.0 10194.1 8390.2 -0.04 -0.04 1.22 5 9.57 15230.6 15231.7 12434.5 -0.02 -0.01 1.22 6 12.76 20287.0 20289.1 16368.0 0.08 0.09 1.24 7 15.95 25327.3 25327.5 20303.3 0.08 0.08 1.25 8 21.27 33720.5 33719.5 26980.5 0.06 0.06 1.25 9 26.59 42089.2 42090.4 33597.9 -0.01 -0.01 1.25 10 31.91 50402.6 50403.1 38969.7 -0.17 -0.16 1.29 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2309 to 64573 e- Average count rate measured: 1578 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.28 +/- 0.01 e-/ADU RON = 5.99 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.083 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay