*** MIDAS logfile (05SEPpl1.2) *** We, 27 Sep 2006 18:15:25 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 27 Sep 2006 18:16:20 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060927 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 275.77 stddev = 4.39 FIERA temperature = 288.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060927 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1773.3 1774.4 1448.5 -0.05 -0.04 1.22 2 2.13 3439.6 3439.6 2946.7 0.20 0.19 1.17 3 4.25 6733.0 6731.9 5502.9 -0.17 -0.18 1.22 4 6.38 10047.0 10046.8 8071.5 -0.11 -0.10 1.24 5 9.57 15014.9 15013.8 12075.7 -0.06 -0.07 1.24 6 12.76 19999.7 20001.5 15914.9 0.04 0.04 1.26 7 15.95 24977.0 24978.5 20025.8 0.07 0.07 1.25 8 21.27 33270.4 33272.4 27055.2 0.09 0.09 1.23 9 26.59 41544.3 41542.6 32385.0 0.06 0.05 1.28 10 31.91 49772.5 49772.5 38160.6 -0.06 -0.06 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2295 to 64397 e- Average count rate measured: 1557 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.29 +/- 0.02 e-/ADU RON = 5.68 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.080 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay