*** MIDAS logfile (05SEPpl1.2) *** We, 13 Sep 2006 16:29:26 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 13 Sep 2006 16:30:21 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060913 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 274.33 stddev = 5.15 FIERA temperature = 287.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060913 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1774.1 1774.1 1437.3 -0.04 0.02 1.23 2 2.13 3428.5 3426.7 2946.9 0.15 0.15 1.16 3 4.25 6701.9 6701.4 5408.7 -0.18 -0.18 1.24 4 6.38 9995.8 9994.7 8063.1 -0.12 -0.12 1.24 5 9.57 14933.6 14933.8 12058.5 -0.07 -0.07 1.24 6 12.76 19891.4 19894.4 15799.2 0.04 0.04 1.26 7 15.95 24839.9 24839.8 19890.4 0.08 0.07 1.25 8 21.27 33084.9 33086.2 26263.9 0.10 0.09 1.26 9 26.59 41313.5 41312.7 32491.9 0.08 0.06 1.27 10 31.91 49489.3 49488.9 38094.8 -0.04 -0.06 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2292 to 63943 e- Average count rate measured: 1547 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.29 +/- 0.01 e-/ADU RON = 6.66 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.086 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay