*** MIDAS logfile (05SEPpl1.2) *** We, 06 Sep 2006 15:55:01 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 06 Sep 2006 15:55:49 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060906 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 275.36 stddev = 5.13 FIERA temperature = 288.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060906 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1697.6 1697.7 1418.0 0.02 -0.01 1.20 2 2.12 3338.6 3337.8 2708.2 0.14 0.14 1.23 3 4.25 6590.3 6590.6 5584.1 -0.20 -0.19 1.18 4 6.38 9863.0 9862.0 7975.9 -0.12 -0.12 1.24 5 9.57 14769.7 14769.1 11796.6 -0.07 -0.07 1.25 6 12.76 19694.9 19694.1 15743.3 0.04 0.05 1.25 7 15.95 24609.0 24610.2 19624.2 0.07 0.08 1.25 8 21.27 32799.3 32800.2 25979.7 0.09 0.10 1.26 9 26.59 40972.3 40972.0 31929.6 0.06 0.07 1.28 10 31.91 49098.3 49099.4 37972.4 -0.05 -0.04 1.29 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2197 to 63545 e- Average count rate measured: 1537 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.29 +/- 0.01 e-/ADU RON = 6.64 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.044 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay