*** MIDAS logfile (05SEPpl1.2) *** Th, 24 Aug 2006 20:41:16 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 24 Aug 2006 20:42:17 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060824 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 275.63 stddev = 5.81 FIERA temperature = 288.3 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060824 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1783.1 1782.0 1513.0 -0.03 -0.03 1.18 2 2.12 3452.0 3451.5 2897.3 0.17 0.17 1.19 3 4.25 6756.3 6755.0 5442.1 -0.17 -0.19 1.24 4 6.38 10080.0 10080.4 8464.0 -0.12 -0.11 1.19 5 9.57 15064.4 15063.3 12322.1 -0.07 -0.07 1.22 6 12.76 20069.8 20069.2 15986.2 0.05 0.05 1.26 7 15.95 25060.4 25061.3 20199.7 0.07 0.07 1.24 8 21.27 33383.2 33383.9 27001.5 0.09 0.10 1.24 9 26.59 41682.2 41683.2 32747.8 0.06 0.06 1.27 10 31.91 49935.1 49937.2 38392.1 -0.06 -0.05 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2303 to 64506 e- Average count rate measured: 1562 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.29 +/- 0.01 e-/ADU RON = 7.51 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.082 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay