*** MIDAS logfile (05SEPpl1.2) *** Su, 20 Aug 2006 18:44:14 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 20 Aug 2006 18:45:12 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060820 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 276.32 stddev = 5.27 FIERA temperature = 288.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060820 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1710.3 1710.8 1523.7 -0.04 -0.05 1.12 2 2.13 3360.2 3361.0 2719.1 0.14 0.16 1.24 3 4.25 6629.5 6630.2 5449.2 -0.19 -0.18 1.22 4 6.38 9918.2 9917.8 8021.4 -0.11 -0.12 1.24 5 9.57 14850.0 14849.2 11878.9 -0.07 -0.07 1.25 6 12.76 19801.9 19801.1 15743.1 0.05 0.05 1.26 7 15.96 24742.1 24740.1 19680.7 0.08 0.08 1.26 8 21.27 32973.5 32972.9 26654.4 0.10 0.10 1.24 9 26.59 41187.7 41187.2 33675.0 0.07 0.07 1.22 10 31.91 49353.3 49355.4 37834.7 -0.04 -0.04 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2179 to 62861 e- Average count rate measured: 1545 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.27 +/- 0.02 e-/ADU RON = 6.71 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.047 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay