*** MIDAS logfile (05SEPpl1.2) *** Th, 10 Aug 2006 14:27:52 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 10 Aug 2006 14:28:47 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060810 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 273.66 stddev = 5.11 FIERA temperature = 287.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060810 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1708.7 1708.2 1383.3 -0.03 -0.03 1.24 2 2.12 3357.9 3357.1 2710.0 0.15 0.14 1.24 3 4.25 6625.7 6624.4 5346.2 -0.19 -0.20 1.24 4 6.38 9913.4 9912.7 7998.4 -0.12 -0.12 1.24 5 9.57 14841.7 14841.9 11894.5 -0.07 -0.07 1.25 6 12.76 19792.9 19791.6 15880.7 0.05 0.05 1.25 7 15.95 24727.4 24726.3 19590.2 0.07 0.07 1.26 8 21.27 32958.3 32957.5 26043.6 0.10 0.10 1.27 9 26.59 41167.7 41168.4 32206.2 0.07 0.08 1.28 10 31.91 49330.9 49333.7 37681.6 -0.04 -0.03 1.31 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2220 to 64111 e- Average count rate measured: 1544 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.30 +/- 0.01 e-/ADU RON = 6.64 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.047 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay