*** MIDAS logfile (05SEPpl1.2) *** Sa, 05 Aug 2006 18:41:28 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 05 Aug 2006 18:42:22 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060805 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 274.21 stddev = 5.26 FIERA temperature = 288.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060805 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1790.0 1789.4 1466.1 0.01 -0.01 1.22 2 2.12 3456.2 3455.5 2829.4 0.16 0.15 1.22 3 4.25 6757.3 6756.6 5437.2 -0.18 -0.19 1.24 4 6.38 10078.7 10077.1 8108.1 -0.12 -0.12 1.24 5 9.57 15056.9 15057.2 12041.2 -0.08 -0.07 1.25 6 12.76 20057.8 20057.1 15982.5 0.04 0.05 1.25 7 15.95 25043.6 25042.9 19917.5 0.07 0.07 1.26 8 21.27 33356.2 33358.5 26682.9 0.09 0.10 1.25 9 26.59 41650.8 41649.2 32275.2 0.06 0.07 1.29 10 31.91 49892.6 49894.2 39049.4 -0.06 -0.05 1.28 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2296 to 63999 e- Average count rate measured: 1560 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.28 +/- 0.01 e-/ADU RON = 6.74 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.087 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay