*** MIDAS logfile (05SEPpl1.2) *** Th, 20 Jul 2006 18:28:32 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 20 Jul 2006 18:29:43 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060720 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 273.16 stddev = 6.03 FIERA temperature = 288.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060720 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1777.8 1779.9 1484.3 0.09 0.04 1.20 2 2.13 3436.2 3438.1 3115.5 -0.04 0.07 1.10 3 4.26 6735.1 6737.1 5515.3 -0.16 -0.09 1.22 4 6.38 10049.7 10047.5 8031.8 -0.03 -0.12 1.25 5 9.57 15016.4 15012.9 12040.9 0.01 -0.06 1.25 6 12.77 19991.7 20005.9 16043.8 0.01 0.09 1.25 7 15.96 24967.7 24965.4 19985.6 0.05 0.02 1.25 8 21.27 33265.5 33263.0 27276.5 0.11 0.08 1.22 9 26.60 41524.2 41530.3 33377.3 0.04 0.04 1.24 10 31.92 49748.4 49759.9 38077.1 -0.08 -0.07 1.31 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2279 to 63749 e- Average count rate measured: 1556 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.28 +/- 0.02 e-/ADU RON = 7.72 +/- 0.12 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.078 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay