*** MIDAS logfile (05SEPpl1.2) *** Tu, 04 Jul 2006 17:11:28 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Tu, 04 Jul 2006 17:12:35 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060704 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 274.46 stddev = 5.23 FIERA temperature = 288.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060704 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1792.0 1789.5 1464.8 0.01 -0.03 1.22 2 2.13 3471.0 3470.1 2797.9 0.15 0.15 1.24 3 4.25 6800.6 6799.1 5456.2 -0.18 -0.17 1.25 4 6.38 10147.7 10148.4 8105.9 -0.12 -0.11 1.25 5 9.57 15168.1 15167.7 12429.6 -0.07 -0.07 1.22 6 12.76 20209.6 20209.7 16043.7 0.05 0.05 1.26 7 15.96 25240.3 25239.8 20140.0 0.08 0.08 1.25 8 21.28 33620.6 33620.7 26770.9 0.09 0.10 1.26 9 26.59 41981.1 41983.8 33069.0 0.06 0.07 1.27 10 31.91 50293.3 50295.0 38683.2 -0.06 -0.05 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2309 to 64844 e- Average count rate measured: 1573 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.29 +/- 0.01 e-/ADU RON = 6.74 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.077 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay