*** MIDAS logfile (05SEPpl1.2) *** We, 21 Jun 2006 19:02:58 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 21 Jun 2006 19:03:55 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060621 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 273.93 stddev = 5.27 FIERA temperature = 288.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060621 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1750.7 1751.8 1459.4 0.00 -0.04 1.20 2 2.13 3423.2 3422.2 2760.9 0.16 0.13 1.24 3 4.25 6736.3 6734.7 5446.2 -0.17 -0.18 1.24 4 6.38 10066.0 10066.1 8053.7 -0.12 -0.12 1.25 5 9.57 15062.2 15061.6 12071.1 -0.07 -0.06 1.25 6 12.76 20078.7 20078.1 16623.2 0.05 0.06 1.21 7 15.95 25081.4 25081.0 19896.1 0.07 0.08 1.26 8 21.28 33419.8 33420.3 26493.1 0.09 0.10 1.26 9 26.59 41738.2 41740.5 32500.4 0.05 0.07 1.28 10 31.91 50007.9 50011.5 38411.9 -0.07 -0.05 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2272 to 64868 e- Average count rate measured: 1565 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.30 +/- 0.01 e-/ADU RON = 6.83 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.058 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay