*** MIDAS logfile (05SEPpl1.2) *** Su, 14 May 2006 16:40:07 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 14 May 2006 16:41:03 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060514 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 325.68 stddev = 4.44 FIERA temperature = 289.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060514 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1847.3 1847.0 1530.9 -0.05 -0.07 1.21 2 2.13 3583.8 3582.8 2927.3 0.18 0.16 1.22 3 4.25 7026.8 7027.6 5978.0 -0.09 -0.07 1.18 4 6.38 10488.1 10488.0 8674.3 -0.03 -0.02 1.21 5 9.57 15672.9 15674.1 12758.8 -0.01 -0.00 1.23 6 12.76 20878.9 20878.4 16994.1 0.09 0.09 1.23 7 15.95 26064.5 26065.4 21043.1 0.08 0.08 1.24 8 21.27 34705.4 34705.2 27680.6 0.05 0.06 1.25 9 26.59 43312.6 43315.2 33859.8 -0.03 -0.02 1.28 10 31.91 51865.0 51866.1 40344.3 -0.20 -0.19 1.29 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2381 to 66845 e- Average count rate measured: 1624 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.29 +/- 0.01 e-/ADU RON = 5.72 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.077 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay