*** MIDAS logfile (05SEPpl1.2) *** Mo, 08 May 2006 22:49:03 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 08 May 2006 22:50:02 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060508 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 325.86 stddev = 5.78 FIERA temperature = 289.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060508 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1848.8 1848.2 1526.2 -0.06 -0.08 1.21 2 2.13 3586.9 3587.4 2929.6 0.16 0.17 1.22 3 4.25 7034.8 7034.7 5765.3 -0.08 -0.08 1.22 4 6.38 10499.9 10499.9 8532.5 -0.02 -0.02 1.23 5 9.57 15689.0 15689.5 12654.3 -0.02 -0.01 1.24 6 12.76 20901.1 20902.1 17342.9 0.09 0.09 1.21 7 15.96 26094.3 26094.5 20999.8 0.09 0.09 1.24 8 21.28 34744.4 34742.3 27814.1 0.06 0.06 1.25 9 26.59 43362.8 43361.8 34162.3 -0.02 -0.03 1.27 10 31.91 51919.8 51920.3 40226.4 -0.20 -0.20 1.29 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2374 to 66676 e- Average count rate measured: 1626 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.28 +/- 0.01 e-/ADU RON = 7.42 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.076 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay